  • Labom CI4100-A1061-F2-H21-S66-M21.1-Y12-T150-K1070-W4022-W1020-W  2002-W2602

CI4100-A1061-F2-H21-S66-M21.1-Y12-T150-K1070-W4022-W1020-W

2002-W2602

PRESSURE TRANSMITTER PASCAL Ci4 with threaded connection

process connection bottom

see data sheet D4-016-1

nominal range: 0...40 bar, Turndown max. 100:1, overload limit 100 bar;

sensor type piezoresistiv

accuracy: <= 0.10% from adjusted measuring span (Lin./Hys./Rep., up to

Turndown <=5:1 no changes)

F2 parameterizing: see below

device ID: LABOM PASCAL Ci4

lower range value: 0

upper range value: + 39,2300

pressure unit: bar

unit of sensor temperature:: °C

damping: 0.0 s

display mode: four values

principle value: pressure

output function: linear

residual current: low < 3,6 mA

HART address 0

output signal: 4...20 mA with HART

Ex design:

TÜV 13 ATEX 120264 X

II 1/2 G, II 2 G Ex ia IIC TX Ga/Gb, Gb

II 1/2 D, II 2 D Ex ia IIIC Txx°C Da/Db, Db

Please pay attention to the application conditions, which can be taken from

the data sheet and the ex-safety instructions XA_010!

high-resolution graphic display with backlight

intuitive 4-button operation, quick access to device data

available languages (English, German), default setting English

material case: stainless steel mat. no. 1.4301/1.4305 (304/303)

material front cover: stainless steel, window non-splittering glass

degree of protection: IP 65 / IP 67

electrical connection: cable gland M 20 x 1.5, polyamide black, for cable

ø 7-13 mm

cable clamps: spring clip terminals up to 1.5 mm2

K1070 process connection: 1/2" NPT

material wetted parts

sensor diaphragm, stainless steel mat. no. 1.4404/1.4435 (316L)

socket, stainless steel mat. no. 1.4404/1.4435 (316L)

marking: stainless steel Tag-plate, fixed with stainless steel wire, marking

see below

Components marked acc. to material inspection certificate

per EN 10204-3.1, wetted parts

CALIBRATION PROTOCOL: with 5 measuring points

 functional safety per EN 61508, classification in SIL2

Write a review

Note: HTML is not translated!
    Bad           Good

Labom CI4100-A1061-F2-H21-S66-M21.1-Y12-T150-K1070-W4022-W1020-W 2002-W2602

  • Product Code: CI4100-A1061-F2-H21-S66-M21.1-Y12-T150-K1070-W4022-W1020-W
  • Availability: In Stock
  • availability : 3-8 week
  • price : send inquiry
  • $0.00

  • Ex Tax: $0.00

Enquire now for a quick price

Tags: Labom, CI4100-A1061-F2-H21-S66-M21.1-Y12-T150-K1070-W4022-W1020-W, 2002-W2602, Ci4, D4-016-1, 0...40, 0.10%, 0.10%, + 39, 2300, 0.0 s, 3, 6 mA, 4...20, IP 65 / IP 67, M 20 x 1.5, 7-13 mm, 1.5 mm2, 1/2", EN 10204-3.1, EN 61508